메뉴 건너뛰기

Qualitative & Quantitative Analyzer

HPLC+GC-MS Triple analyzer, GC-MS, LC-MS, IC, FTIR, XRD, UV/Vis Spectrometer

ICP1000 Inductively Coupled Plasma - Atomic Emission Spectrometer (ICP-AES)

Features
  • Over 70 elements can be measured
  • Fast analysis, more than 10 elements per minute
  • Excellent detection limits, at ppb level for most elements
  • Wide linear dynamic range, reaching 6 orders of magnitude, from ppb to percentage
Application
  • Rare earth metal
  • Silicon materials, magnetic materials processing industry
  • Metallurgical industry: analysis of impurity elements influencing the quality of metal materials. Water quality control
  • Petrochemical: measurements of more than 30 elements in crude oil
  • Pharmacy, hygiene, agriculture, environmental protection, commodities and food industry
Performance Characteristics
  • Advanced optical design with a fully illuminated holographic grating, with spectral interference correction and high light throughput for improved detection limits
  • Wide wavelength range
  • Computer controlled plasma platform optimizes the viewing position to reduce interferences, improve SNR and minimize background emissions
  • Software controlled flow of carrier gas, plasma gas and auxiliary gas provide real time monitor as well as highly stability of flux
  • Various types of touches, nebulizer as well as spray chamber are optional
  • Sample waste drained by peristaltic pump ensures stable sample flow rate; Fast pump mode improves productivity
  • The brand new robust free-running, 40.68 MHz RF generator that delivers unsurpassed performance
  • Rapid and accurate automatic coupling system ensures the ultimate power transfer efficiency and stability
Specifications
Solid State Power (RF) Frequency 40.68MHz
Frequency stability ≤0.05%
Output power 800W ~ 1200W, continuously adjustable with power efficiency more than 65%
Output power stability <0.3%
Electromagnetic leakage radiation intensity E<2V/m at 30cm distance from equipment
Electronic Measurement & Control Circuit Photomultiplier tube specification R212UH/R928
Negative high voltage -200 ~ -1000 V
Circuit measuring range 10-9 ~10-4 A
Signal acquisition VF conversion 1mV ? 100Hz
Inlet system Automizer concentric atomizer
Induction coil ID 25mm x 3, equipped with three concentric quartz torch tubes of ED 20 mm (ID: internal diameter; ED: external diameter)
Argon for carrier gas, auxiliary gas and Plasma gas
Carrier gas: 0-5L/min (control by system)
Auxiliary gas: 10-100L/h or 0.16-1.66L/min
Plasma Gas: 100-1000L/h or 1.6-16 L/min
Monochromator Optical type Czerny turner
Focal length 1000mm
Resolution <0.005nm
Grating specification Huge holographic rating with 3600L/mil or 2400 L/mil and 80mm x 110mm of ruling area
Wavelength range 195-500nm for 3600 line grating
195-800 nm for 2400 line grating
Technical Specifications Suitable sample content range Liquid sample: 0.01ppm~ several thousands ppm
Solid or powder sample: 0.001%~70%
Repeatability short-term stability, RSD≤1.5%
Long-term stability: RSD<2%
Test speed 10 elements/min
Limits of detection (LOD, μg/L) for typical elements 1ppb ~ 10ppb
 
ICP1000 Inductively Coupled Plasma Atomic Emission(ICP-AES) Spectrometer

 
share