메뉴 건너뛰기

Application

가속속도열량계

Heat Losses and Errors in Accelerating Rate Calorimetry

YEONJIN 2009-01-08 17:42:42 6914

Heat losses in the Accelerating Rate Calorimeter may be divided into two types. Heat lost into the bomb (the phi correction) and heat lost from the bomb. The latter, the φ correction is discussed in detail in THT Technical Information Sheets No 22, 23, 24.

In the Accelerating Rate Calorimeter, the φ correction is the only correction necessary to carried out with the sample data to get excellent adiabatic data. In experiments a full simulation of the potentially undesired chemical reaction is carried out. When this is done under conditions which do not lead to self-heating rates so high that the accurate tracking or following of the bomb/sample by the calorimeter is lost there are no errors in the data. There is a limit of maximum heating of the calorimeter and of course if the self-heating of the sample/bomb exceed this, then there is a divergence in temperature and the possibility of heat loss from the sample/bomb system. Ie a loss of
operational adiabaticity.


전문은 첨부된 "기초자료"를 다운로드 하셔서 열람하세요.

Reference:
ARC Technical Information Sheet No. 21, Thermal Hazard Technology, UK

검색 결과 31건

  • 분류 : 가속속도열량계