Liquid Analysis Systems사의 L450 series 화학 분석기는 pH, ORP, ISE, 탁도계 및 다양한 센서를 사용하여 많은 수성 화학 종의 온라인 습식 화학 적정법에 따라 분석합니다. L450의 기능은 신뢰성 및 사용 편의성, 최소 운영 비용에 최적화되어 있습니다.
L450 Series Features
• Auto sample retrieval and preparation
• Result replication and range check
• Auto-check of titrant, sample, and sensors
• Output options for recording, alarming, and replenishing.
Through its display and keypad, users can view process status and history, and modify analysis intervals and other configuration parameters. This series offers a variety of hardware options for stream selection, sample preparation, and sample/reagent delivery.
Series Options
• Multi-stream and multi-parameter analysis
• Grab sample port
• Standard precision titrant dose pumps with better than ±3% accuracy
• High precision titrant burette pumps with better than ±1% accuracy
• Auto sample-filter back flush
• Windows™ PC software for process oversight and control
• Replenishment systems
L450Zn At Line Zinc Titrator
The L450Zn at line nickel titrator measures zinc ion concentration in water, wastewater, or process streams by means of a robust EDTA titration. Its differential redox endpoint detection method eliminates the need for sensor calibration or replacement. Because peristaltic pumps and colorimetric dyes and indicators are not utilized, maintenance and reagent costs are minimized.
L450Zn At Line Zinc Titrator has configuration of Base model with one standard precision titrant dose pumps, single stream inlet, and flow-thru sample capture.
Options
- AE Eductor pump option. Used in place of electric pump.
- GS Grab sample capability. Sipper tube inlet.
- HP High precision. 0.1 µl resolution titration burette.
- MSx Multi-stream, x = number of streams.
- WF Auto water back flush of sample filter. (Use with -EP.)
- WM Windows™ Process Manager PC software
Specifications
Ranges | Configurable over 20 to 1 range |
Accuracy | Standard method: ±3% of range (Accuracy is configurable and is inversely proportional to reagent consumption.) |
Method | EDTA titration after pH adjustment |
Cycle time | 7 minutes, typical |
Drift | ≤ 1% / yr |
Reagent consumption | Standard precision model: 0.2 to 1 ml / test |
Power | 100/240 VAC 50/60 Hz |
Sample pressure | 0 to 25 psig |
Sample temperature | 10 to 50 °C |
Sample connection | ¼” tube |
Sample streams | Standard:1 Optional: up to 8 |
Water (only used with -SP or -WP options) | 100K ohm resistivity, minimum 20 to 60 psi, ¼” NPT-F connection |
Air supply | 65 psi, oil-free, ¼” NPT-F connection |
Drain | ½” tube connection |
Vent (to drain) | ¼” tube connection |
Enclosure | 20” H x 16” W x 13” D, wall mount, NEMA 4X/IP66 |
Display/Touchscreen | 3” x 2.2”, LCD with backlight Relay output Form C relay standard. Other per user requirement. |
Analog output option | 4-20 mA. Other per user requirement. |
Network output option | Ethernet: email of status and alarms |
Specifications subject to change