HCS601GXY-IRM Heating & Cooling Stage
- 
	
Flagship stage for microscopy FTIR
 - 
	
-190°C - 600°C with liquid nitrogen cooling option
 - 
	
24 mm × 24 mm sample area with wide angle viewing aperture
 - 
	
XY sample positioning
 - 
	
Sealed chamber with gas purging capability
 
Thermal Features
| 
			 Temperature Control  | 
			
			 mK2000 with programmable precision switching PID method  | 
		
|---|---|
| 
			 Thermal Block  | 
			
			 Silver for superior thermal uniformity  | 
		
| 
			 Temperature Minimum  | 
			
			 -190°C (with optional liquid N2 cooling)  | 
		
| 
			 Temperature Maximum  | 
			
			 600°C  | 
		
| 
			 Temperature Sensor  | 
			
			 100 Ω Platinum RTD  | 
		
| 
			 Maximum Heating Rate  | 
			
			 +200°C per minute at 100°C  | 
		
| 
			 Maximum Cooling Rate  | 
			
			 -50°C per minute at 100°C  | 
		
| 
			 Minimum Heating and Cooling Rate  | 
			
			 ±0.01°C per minute  | 
		
| 
			 Temperature Resolution  | 
			
			 0.01°C  | 
		
| 
			 Temperature Stability  | 
			
			 ±0.05°C (above 25°C), ±0.1°C (below 25°C)  | 
		
| 
			 Power Supply  | 
			
			 Universal power input  | 
		
| 
			 Software  | 
			
			 Windows software to record and export temperature-time data  | 
		
Optical Features
| 
			 Optical access  | 
			
			 Reflection and transmission capability  | 
		
|---|---|
| 
			 Optical windows  | 
			
			 Removable and exchangeable windows permit full-spectrum transparency  | 
		
| 
			 Minimum Objective Working Distance  | 
			
			 10.5 mm  | 
		
| 
			 Minimum Condenser Working Distance  | 
			
			 10 mm  | 
		
| 
			 Top Window  | 
			
			 27 mm diameter  | 
		
| 
			 Top Viewing Angle  | 
			
			 ±57.5° from normal  | 
		
| 
			 Transmission Aperture  | 
			
			 2 mm diameter  | 
		
| 
			 Bottom Window  | 
			
			 27 mm diameter  | 
		
| 
			 Bottom Viewing Angle  | 
			
			 ±55.7° from normal  | 
		
| 
			 Window Defrost  | 
			
			 Integrated external window defrost  | 
		
Structural Features
| 
			 Sample Area  | 
			
			 24 mm x 24 mm  | 
		
|---|---|
| 
			 Chamber Height  | 
			
			 6.5 mm without removable inner cover 2.5 mm with removable inner cover  | 
		
| 
			 Sample Access  | 
			
			 Side-loading cover for quick sample access without refocusing instruments  | 
		
| 
			 Sample Positioning  | 
			
			 10 mm fine travel with Vernier XY dials for remote manipulation in sealed chambers  | 
		
| 
			 Atmosphere Control  | 
			
			 Sealed chamber with gas purging capability to control humidity, condensation, and oxidation  | 
		
| 
			 Frame Cooling  | 
			
			 Integrated frame cooling via hose barb ports with optional chiller system  | 
		
| 
			 Mounting  | 
			
			 Horizontal and vertical mounting capability  | 
		
| 
			 Frame Dimensions  | 
			
			 180 mm x 100 mm x 20.5 mm  | 
		
| 
			 Weight  | 
			
			 1200 g light-weight aluminum body  | 
		