L450 Chloride Analyzer 온라인 염소측정기 (염화이온, Cl-)


The online L450 Chloride analyzer measures chloride in water by means of a selective and accurate titration with silver or mercuric nitrate. With the differential endpoint detection method, the internal Pt electrode does not require calibration or replacement, and electrode depletion and drift common with other methods is eliminated. No colorimetric dyes or indicators are required. For dependable process control, each analysis is automatically replicated and/or trend checked prior to posting, alarming, or replenishing. Sampling can be drawn from pressurized sample stream or pumped from overflow sampler. For sample with particulate, a sample strainer with automatic backflush is offered.

L450Chloride Analyzer has configuration of Base model with standard precision (20 µl) titrant dose pumps, one sample/waste pump, single stream inlet, 4-20 mA output.

Liquid Analysis Systems사의 L450 series 화학 분석기는 pH, ORP, ISE, 탁도계 및 다양한 센서를 사용하여 많은 수성 화학 종의 온라인 습식 화학 적정법에 따라 분석합니다. L450의 기능은 신뢰성 및 사용 편의성, 최소 운영 비용에 최적화되어 있습니다.


L450 Series Features

• Auto sample retrieval and preparation

• Result range and trend check

• Auto check of titrant, sample, and electrodes

• Output options for recording, alarming, dosing, and/or replenishing

Through its display and keypad, users can view process status and history, and modify analysis intervals and other configuration parameters. This series offers a variety of hardware options for stream selection, sample preparation, and sample/reagent delivery.

 

Series Options

• Multi-stream and multi-parameter analysis

• Standard precision titrant dose pumps with better than ±1.5% accuracy

• Grab sample port

• Replenishment and dosing systems

• RS 485/Ethernet networking/E mail messaging

 

L450 Chloride Analyzer

The online L450 Chloride analyzer measures chloride in water by means of a selective and accurate titration with silver or mercuric nitrate. With the differential endpoint detection method, the internal Pt electrode does not require calibration or replacement, and electrode depletion and drift common with other methods is eliminated. No colorimetric dyes or indicators are required. For dependable process control, each analysis is automatically replicated and/or trend checked prior to posting, alarming, or replenishing. Sampling can be drawn from pressurized sample stream or pumped from overflow sampler. For sample with particulate, a sample strainer with automatic backflush is offered.

L450Chloride Analyzer has configuration of Base model with standard precision (20 µl) titrant dose pumps, one sample/waste pump, single stream inlet, 4-20 mA output.

 

Options

DI Digital input for remote control of analyses or replenishment

DO Digital output relay with settable trip point or other control function

GS1 Auto grab sample with sipper tube inlet.

HP High precision titrant pump

MSx Multi-stream sampling, x = number of streams.

RRx Replenishment control (Relay or Pneumatic output options)

SK Spares kit (basic and extended versions available)

SP Sample pump. For unpressurized samples.

 

Specifications

Method Differential titration with silver nitrate and Pt electrode. Mercuric nitrate titrant option for ranges under 2.5 ppm. pH buffering is available for strongly alkaline sample.
Ranges

Configurable over ≤50:1 range, e.g., 20 to 1000 ppm Cl

Repeatability Settable: 2% of range, typical at 2 replicates. Setting drives number of replicates.
Interferences Cyanide, bromide, iodide, orthophosphate
Cycle time Minimum 2 to 10 minutes per stream
Stability Better than 1% / yr, excluding titrant stability.
Reagent consumption ~200 µl/test, dependant on process value and range
Power required 100/240 VAC or 24 VDC
Sample streams Standard:1 Optional: up to 4
Sample conditions 5 to 25 psig, 15 to 60 °C
Sample connection ¼” NPT
Drain ½” NPT. Drain must be vented or unpressurized.
Enclosure 24” H x 20” W x 10” D, wall mount, NEMA 4X/IP66
Display/Touchscreen 3” x 2.2”, LCD with backlight
Outputs

Standard: 4-20 mA. Maintenance alarm relay.

Optional: process alarm relay with settable trip point, Ethernet, and/or serial per user requirement

Specifications are subject to review of sample conditions.

L450 Chloride Analyzer 온라인 염소측정기 (염화이온, Cl-)

The online L450 Chloride analyzer measures chloride in water by means of a selective and accurate titration with silver or mercuric nitrate. With the differential endpoint detection method, the internal Pt electrode does not require calibration or replacement, and electrode depletion and drift common with other methods is eliminated. No colorimetric dyes or indicators are required. For dependable process control, each analysis is automatically replicated and/or trend checked prior to posting, alarming, or replenishing. Sampling can be drawn from pressurized sample stream or pumped from overflow sampler. For sample with particulate, a sample strainer with automatic backflush is offered.

L450Chloride Analyzer has configuration of Base model with standard precision (20 µl) titrant dose pumps, one sample/waste pump, single stream inlet, 4-20 mA output.


 

Liquid Analysis Systems, Inc. USA
share
« 1 2 3 4 »
« 3 / 4 »