메뉴 건너뛰기

Thermal Properties Analyzer

Thermal Conductivity (Lambda), Ash Fusion, Thermal Screening, Microscopic Hot/Cold Stage, Wafer Chuck, Thermal Plate, Thermal Probe Station

HCP600G-CAP Specialty Thermal Plate


  • Compact Hot/Cold Plate for capillary tube applications
  • Manipulate capillary tube while maintaining chamber atmosphere
  • -190°C - 600°C with liquid nitrogen cooling option
  • 28 mm × 30 mm sample area
  • Gas-tight chamber

 

Description

The HCP600G-CAP gas tight plate is designed for capillary tube applications where both thermal and atmospheric control is critical. Using a silver heating and cooling block, this plate provides a wide temperature range with precise thermal uniformity. The gas tight chamber provides a closed chamber to eliminate oxidation, aid in humidity studies, or conserve expensive reaction gasses.

 

Thermal Features

Temperature Control

mK2000 with programmable precision switching PID method

Thermal Block

Silver

Sample Thermal Cover

Optional removable Inner sample cover with additional window

Temperature Minimum

-190°C (with optional liquid N2 cooling)

Temperature Maximum

600°C

Temperature Sensor

100 Ω Platinum RTD

Maximum Heating Rate

+150°C per minute at 100°C

Maximum Cooling Rate

-50°C per minute at 100°C

Minimum Heating and Cooling Rate

±0.01°C per minute

Temperature Resolution

0.01°C

Temperature Stability

±0.05°C (above 25°C), ±0.1°C (below 25°C)

Power supply

Universal power input

Software

Windows compatible software to record and export temperature-time data; remotely control system via VCP connection (Virtual Comm Port)

 

Optical Features

Optical access

Reflection only

Optical windows

Removable and exchangeable windows

Minimum Objective Working Distance

5 mm

Top Aperture

27 mm diameter

Window Defrost

Integrated external window defrost

 

Sturctural Features

Sample Area

28 mm x 30 mm

Chamber Height

2.8 mm

Sample Access

Side-loading capillary tube for quick sample access without refocusing  instruments

Atmosphere Control

Sealed chamber with gas purging capability to control humidity,  condensation, and oxidation

Frame Cooling

Integrated frame cooling with optional chiller system

Mounting

Horizontal capability

 

 

 


 

Instec, Inc. USA
share