SDTGA6000a Thermogravimetric Analyzer (Proximate Analyzer)

Highights

  • 19 samples can be loaded once + 1 blank sample position for analyzing the moisture, ash and volatile matter.
  • Unique sample weighing mode, single sample or batch sample weighing are both available, makes the operation convenient.
  • Automatic temperature calibration function is available.
  • Flexible software design, test procedure, analysis time, temperature etc. can be customized.
  • LED displays on furnaces to show real-time balance weight data, it makes the sample weighing more convenient.
  • Single and dual furnace configurations available.

 

Specification

Description

SDTGA6000A

Test index

Moisture, ash ,volatile matter and  LOI

Continuous test in batches

Yes

Standards

ASTM D3173/D3174/D3175/DD7582, ISO 17246, ISO 18123/18122/18134, ISO 11722, ISO 1171

Precision

0.02% RSD (1σ, 1g sample)

Furnace temperature

Room temperature ~ 1000℃

Temperature control precision

±2℃

Temperature resolution

0.1℃

Reproducibility

  • Ash: 0.30% (<15.00%); 0.50% (15.00~30.00%); 0.70% (>30.00%)

  • Volatile matter: 0.50% (<20.00%); 1.00% (20.00~40.00%); 1.50% (>40.00%)

Cooling time required for batches of test

No

Furnace Cooling (forced air/N2)

950 to 500°C <30 min

Balance Resolution

0.0001g

Sample introduction and taking

Automatic

Max. Samples

19 samples + 1 blank sample position

Quantity of sample / test time:

Full indexes for 19 samples< 120 min (Quick method)

Sample weight

0.5 up to 5g

Max. power

4 kW

Interface:

RS232

Furnace Heating

Resistance Wound

Temperature Calibration

Standard thermocouple

Temperature control precision

±2℃ for moisture
±5℃ for ash, volatile matter

Gas Flow Rate (ASTM)

  • Moisture:Nitrogen or Air 2~4(Furnace Volume /min)

  • Volatile matter:Nitrogen 2~4(Furnace Volume /min)

Power requirement

AC220V ±10%,50 / 60Hz

Gas requirement

Nitrogen/Oxygen/Compressed Air

Gas Pressure

Air 5 to 6 bar (75 to 90 psi)

Minimum Gas Purity

Air/Nitrogen/Oxygen 99.5%; dry, oil-free

Sample weighing

Under normal temp.

Test Temperature

  • Moisture 105℃,

  • Ash 815℃,

  • Volatile matter 920℃~950℃ 

Can be customized

Humidity

<80%

Dimensions

550 x 590 x 530mm

Weight

70kg

Operational Control

Windows7/ Windows10 on external PC

Advance Calculation

Yes

Analysis Report

Yes

Statistics Report

Yes

System Monitor

Yes


 