SDTGA6000은 석탄, 코크스, 소성 석유 코크스 및 다양한  고체 바이오매스 연료의 수분과 회분, 휘발분 함량을 측정하고 고정탄소 및 발열량을 계산하는 데 사용할 수 있습니다.

  • 듀얼 퍼니스 구조.
  • 안정성 있는 성능과 정확한 측정결과를 보장합니다.
  • 수분 및 회분에 대한 19개 sample과 1개의 blank.
  • 휘발성 물질에 대한 25개 sample과 + 1개의 blank.

 

Highlights

  1. Dual furnaces structure, the system consists of two parts, one for moisture and ash analysis, other for volatile matter analysis(ASTM only.
  2. For moisture and ash analysis, 19 samples can be loaded once + 1 blank sample position;
  3. For volatile matter analysis, vertical furnace structure is adopted with 26 positions. Upper part for high temperature volatile matter test while lower part for low temperature cooling and weighing, makes the test result more accurate and test efficiency greatly improved.
  4. Flexible software design, test procedure, analysis time, temperature etc. can be customized;
  5. Unique sample weighing mode, single sample weighing and batch sample weighing are both available, makes the operation convenient.
  6. After finishing the moisture and ash test, furnace cover will be opened automatically for furnace cooling down, thus to shorten the waiting time for next batch of test, makes the operation safe and efficient;
  7. LED displays on both furnaces to show real-time balance weight data, it makes the sample weighing more convenient.
  8. Automatic temperature calibration function is available.
  9. Unique design to ensure the stability of heating element.
  10. High Stability and Reliability.
  11. Patented technology of heat-resistant to ensure stability and reliability.
  12. Built-in balance ensures unbeatable accuracy, comply with thermogravimetric method.
  13. Using blank crucible to calibrate the influence of the thermal buoyancy.
  14. LED displays on both furnaces to show real-time balance weight data, it makes the sample weighing more convenient.
  15. Automatic temperature calibration function is available.


Specification

Description

SDTGA6000 (include SDTGA6000A and SDTGA6000V)

Test index

Moisture, ash ,volatile matter and  LOI

Continuous test in batches

Yes

Standards

ISO 17246, ISO 18123/18122/18134, ISO 11722, ISO 1171, ISO 562, ASTM D3173/D3174/D3175/D7582, GB/T 30732, GB/T 212

Precision

0.02% RSD

Furnace

 Dual furnaces, one for moisture and ash, one for volatile matter

Furnace temperature

Room temperature~1000℃

Temperature control precision

±2℃

Temperature resolution

0.1℃

Reproducibility

  • Ash: 0.30% (<15.00%); 0.50% (15.00~30.00%); 0.70% (>30.00%)
  • Volatile matter: 0.50% (<20.00%); 1.00% (20.00~40.00%); 1.50% (>40.00%)

Cooling time required for batches of test

No

Furnace Cooling (forced air/N2)

950 to 500°C<30 min

Balance Resolution

0.0001g

Sample introduction and taking

Automatic

Max. Samples

19 samples + 1 blank sample position (Moisture & Ash)

25 samples + 1 blank sample position (Volatile matter)

Quantity of sample / test time:

Full indexes for 19 samples< 120 min (Quick method)

Sample weight

0.5 up to 5g

Max. power

4kW for moisture/ash testing;

2.5kW for volatile matter testing

Interface:

RS232

Furnace Heating

Resistance Wound

Temperature Calibration

Standard thermocouple

Temp. control precision

±2℃ (moisture) ±5℃ (ash, volatile matter)

Gas Flow Rate

  • Moisture:Nitrogen or Air 2~4(Furnace Volume /min)
  • Volatile matter:Nitrogen 2~4(Furnace Volume /min)

Power requirement

AC 220 V±10%,50 / 60Hz

Gas requirement

Nitrogen/Oxygen/Compressed Air

Gas Pressure

Air 5 to 6 bar (75 to 90 psi)

Minimum Gas Purity

Air/Nitrogen/Oxygen 99.5%; dry, oil-free

Sample weighing

Under normal temp.

Test Temperature

  • Moisture 105℃,
  • Ash 815℃,
  • Volatile matter 920℃~950℃ 
  • Can be customized

Humidity

<80%

Dimensions

moisture& ash analyzer: 550*590*530mm      

volatile matter analyzer: 550*650*815mm

Weight

moisture & ash analyzer: 70kg      

volatile matter analyzer: 95kg

Operational Control

Windows®-based software on external PC

Advance Calculation

Yes

Analysis Report

Yes

Statistics Report

Yes

System Monitor

Yes

 

SDTGA6000 Thermogravimetric Analyzer (Proximate Analyzer)

SDTGA6000 can be used to determine the moisture, ash and volatile matter as well as calculation of fixed carbon and calorific value in coal, coke, calcined petrol coke and other solid biomass fuel.

Dual Furnaces,stable performance, accurate test result.
19 samples+1 blank sample for moisture & ash; 25 samples+1 blank sample for volatile matter


 

share
공업분석기 (Auto TGA) SDLOI-1000 Loss on Ignition Analyzer SDLOI-1000 can be used to determine the LOI,moisture and ash of solid samples.Including moisture an
댓글 0
23,059
문의
하기